beautypg.com

Using your meter – Myron L EP-10 User Manual

Page 2

background image

1. Turn the Range Switch to the desired range. if you don’t

know which range to use, set it to 1000.

2. Rinse the cell cup three times with the sample you want

to test. (For very hot or very cold samples see

TEMPERaTuRE CoMPEnSaTion below.)
never fill the cell by dipping the meter into

water!

3. Fill the cell with another sample to at least 1/4” (6mm)

above the upper electrode.

4. Press the black button.

5. Read the dial’s black scale value indicated by the

pointer. if the pointer is very low on the left or goes off

the scale to the right, try the next lower or higher range.

6. Multiply the dial’s black scale value by the range setting

to determine conductivity in micromhos (microsiemens).

7. if the pointer goes off the scale to the right on the x1000

range, use an RE-10 range extender (see

aCCESSoRiES).
note:

When you are finished with the meter, RinSE THE CELL

CuP with clean water, preferably distilled or deionized.
temperature compensation: For very hot or very

cold solutions, let the three rinse samples each remain in

the cell for several seconds. Then immediately fill the cell

with the sample you want to test (step 3 above). This

allows the automatic temperature compensation feature

time to work properly.
resistivity measurement: Model EP only. accurate

measurement of high purity water is difficult even under

controlled conditions. Water temperature and even carbon

dioxide (Co

2

) in the atmosphere will greatly influence

resistivity. use the following

methods for best accuracy:

1. Turn the range switch to the x.1 range (with the “Ω”

symbol around it).

2. Hold your meter close to the high purity water outlet.

While pressing the button, allow the water to flow into

the Cell Cup and out again, continuously purging the

cell. Do not allow air bubbles to collect.

3. as the cell is washed clean of trace contamination, the

reading will drop and eventually stabilize. Read the

resistivity directly in megohms on the red scale. do

not multiply the dial reading by the range switch value.
note:

The closer the tested water temperature is to 25ºC

(77ºF), and the less it is exposed to air, the more accurate

your reading will be.
caution

always rinse the cell cup with pure water immediately

after use or cleaning. This will prevent the build-up of

unwanted deposits and possible contamination of the next

sample.
do not use with samples hotter than specified. The

readings WiLL noT be accurate.
do not splash solvents such as lacquer thinner, acetone,

benzene or chlorinated solvents on the plastic case.
do not fix or modify the meter. That will void your

warranty. See SERviCE for details or consult Myron L

Company
do not dip the instrument into water. if water

does get inside the instrument, see MainTEnanCE for

instructions on drying it.