Drain installation – Nor-Lake Stability Chambers User Manual
Page 3
Duplex/ or European Outlet
- (Optional) is located near the top of the interior cabinet back. This
is a 15 amp 115 volt duplex PN 88010300 with its own power supply cord. Duplex outlet power
cord is wrapped up on the back of the cabinet top. Outlet is not inter-wired thru main cabinet
supply and must be plugged into a 115 volt/ or 220/240V power supply. Locate cabinet within 8
feet of the wall receptacle.
Access Port
- (Optional) is provided with a spring loaded cover on the right hand side of the
cabinet.
HUMIDITY INSTALLATION (if applicable)
The water supply should have a dedicated shutoff valve and regulator installed. Connect the water
supply to the humidifier (the operating pressure range is 10-60 psi). For best results, provide a
source of reverse osmosis water with a resistance of .05 to 1.0mΩ, which will reduce the scale
build up and cleaning frequency. Unit is equipped for water supply with 1/8” Male Pipe Threads.
The humidifier reservoir water level may be adjusted by using the adjusting thumb screw located
on the float valve (see below).
The humidifier is a precision built instrument that,
given proper care, will provide years of
dependable service. Cleaning the humidifier
periodically is all that is required to insure proper
operation. Depending on local water conditions,
mineral solids and other matter may accumulate
in various parts of the unit. These accumulations
must be removed so that water flows freely for
efficient operation. Failure to clean the humidifier
may result in malfunction which will lead to repairs.
- 3 -
If higher humidity levels are unobtainable, loosen the
reducing coupling clamp and remove reducing coupling
boot and reducer PVC.
DRAIN INSTALLATION
The Stability Chamber has a floor drain in the bottom of the
cabinet. A silicone drain stopper has been placed in this drain.
The drain stopper will prevent undesirable liquids from running
into the drain. The stopper can be removed if any possible
liquids flowing into this drain are acceptable to go to the floor drain.