Maintenance, Periodic cleaning, Maintenance service and analysis guide – Nor-Lake Large Capacity CO2 Incubator User Manual
Page 6: Drawer slide removal/adjustment, Figure c
side inward and lift up the drawer. Slide the drawer towards you and remove.
Figure C
Drawer Slide Removal/Adjustment
Each drawer slide is independently removable and can be adjusted to different levels. First remove
the drawer per above instructions. Release the locking tabs located on the drawer slide, move the
front tab up and release the tab in the back of the cabinet, by sliding the locking tab towards the
front. See illustration. Unhook the drawer slide assembly from both the front and back shelf
standards by lifting up on the assembly.
To install a drawer slide, reverse the process used to remove the slide. Secure the locking tabs into
position after the drawer slide assembly is in place. Reinstall the drawer.
MAINTENANCE
Periodic Cleaning
Beginning with the initial installation, the interior surfaces of the cabinet should be periodically wiped
down with a solution of warm water and baking soda. This solution will remove any odors from
spillage that has occurred. The exterior of the cabinet should also be cleaned frequently with a
commercial grade of glass cleaner. Caution: Do not use an abrasive or alkaline solution.
All moving parts have been permanently lubricated and will generally require no maintenance.
MAINTENANCE SERVICE AND ANALYSIS GUIDE
MALFUNCTION
POSSIBLE CAUSE
SOLUTION
No power to optional
1. Internal circuit breaker tripped
1. Reset circuit breaker
duplex receptacle
2. Wiring incorrect
2. Check wiring against the diagram
Heater inoperative
1. High limit thermostat tripped
1. Manually reset thermostat
2. Wiring incorrect
2. Check wiring against diagram
No power to cabinet
1. Service cord unplugged
1. Plug in service cord
2. Circuit breaker supplying
2. Determine reason and correct
main electrical receptacle tripped
3. Wiring incorrect
3. Check wiring against diagram
4. Main cabinet power switch off
4. Turn on power switch
Objectionable noise
1. Vibrating fan blade
1. Replace fan blade
2. Worn fan motor bearings
2. Replace fan motor
CO
2
measurement does not equal
1. Poor analyzer
1. Verify analyzer operation
the display
2. Display not calibrated
2. Check calibration (see control manual)
CO
2
measurement less than
1. CO
2
regulator set too low
1. Adjust regulator (not to exceed 25 PSI)
setpoint
2. Door recent opened or poor seal
2. Correct door seal
3.
CO
2
tank(s) low
3. Change tank(s)
4. Display not calibrated
4. Check calibration (see control manual)
CO
2
and analyzer read zero
1. Loss of CO
2
supply
1. Identify and correct
(setpoint ok)
2. Defective solenoid 2.
Replace
solenoid
3. Supply tank(s) empty
3. Change tank(s)
6
11/10 Rev. B 143314