6.8 TOR – Maximum Reading Time

Timeout for the reader. TOR is used in operation mode 2 as maximum gating time for a

reading process . The length of the maximum gating time results from the equation gat-

ing_time = TOR * TB.

The time constant TB (Time Base) has always the default value 100ms.
Input format:

TOR parameter

Parameter:

Function

0

limits the reading process duration to exactly one reading cycle

1

limits the reading process duration to maximum 1 times TB

2

limits the reading process duration to maximum 2 times TB

...
255

limits the reading process duration to maximum 256 times TB

Output (example):

2

6.9 VS- Show Parameter

The command VS lists all current parameter settings.

Input format:

VS

Output (example):

EC 0
BD 2

MD 2

6.10 VSAVE – Save Parameter

All operating parameters temporarily stored are saved permanently using VSAVE.
Input format:

VSAVE

Output:

ok