background image

Section 4

4-4

Diode Checks

Tests diodes and rectifiers for proper operation.

1) Set RANGE switch to HIGH.

2) Set FUNCTION switch to OHMS.

3) Connect TEST leads as shown in Figure 1.

( YELLOW to anode — BLACK to cathode)

4) The bottom TEST light should be ON. If a

different TEST light is on or all lights are off, the
diode is defective.

5) Reverse test lead connections as shown in

Figure 2.

( YELLOW to cathode — BLACK to anode)

6) The top TEST light should be ON. If a different

TEST light is on or all lights are off, the diode is
defective.

Note: Always make sure all connections are good. If
necessary, scrape away corrosion, paint, etc. at the
contact points.

FUNCTION

VOLTS

OHMS

FREQUENCY

RANGE

HIGH

LOW

POWER OFF

TEST

TEST

Yellow

Black

Fig. 2

Fig. 1

Black

Yellow