AZI Customer Service 800-528-7411 or 602-470-1414

Page 85 of 92

No heat (cont.)

Door sensor failure

Check display for LID OPEN
message when open during a
test, call AZI at 800-528-7411
or 602-470-1414 for
assistance

Temperature setpoint too low

Raise temperature above
room ambient or standby
temperature, see pages 25 &
68

No power to
instrument

Power switch off

Check switch position,
confirm power cord seating,
remove and test fuse, see
page 14, test the outlet with a
lamp or meter.

Power cord disconnected

Fuse blown

Outlet not powered

No weight reading

Balance not initialized

Turn power off, wait 5
seconds, then turn power on,
wait 15 seconds. Call AZI at
800-528-7411 or
602-470-1414 for assistance

Erratic results

Poor AC ground

Ensure instrument is
grounded from case ground
to earth ground, Call AZI at
800-528-7411 or
602-470-1414

Dirt in chamber

Clean unit,
see pages 78 to 82

Dirt in pan support

See page 78 or call AZI at
800-528-7411 or
602-470-1414 for assistance

Display erratic

Unit cold (near freezing)

Allow to warm-up

THE BALANCE
REPORTS
UNDERLOAD…

Unit powered ON without pan
support in place.

Power off instrument; Install
pan support and power on
instrument, see page 13

UNDER appears on
display when
[START] is pressed

Pan support has been
removed since instrument was
powered ON

Power off instrument; Install
pan support and power on
instrument, see page 13

Instrument won't tare Test chamber temperature

fluctuates/changes

Allow system to stabilize.

Check tare options,

see page 30

Dirt in test chamber or under
pan support

Clean instrument, see page
78 to 82 and following