Moistures too high or
solids too low

Temperature set too high

Lower temperature,

see pages 25 & 68

Bent sample pan

Use new pan

Dirty sample pan

Use new pan

Debris in balance gaps

Call AZI at 800-528-7411 or
602-470-1414 for assistance

Dirty heater assembly

Clean heater assembly,

see page 80

Ending criteria incorrect

Determine new ending
criteria, see page 26

Call AZI at 800-528-7411 or
602-470-1414 for assistance

Moistures too low or
solids too high

Temperature too low

Increase temperature,

see pages 25 & 68

Sample size too small or too
large

Adjust sample size,

see pages 29 & 66

Sample needs spreading

Spread sample evenly,

see page 65

Ending criteria incorrect

Determine new ending
criteria, see page 26

Call AZI at 800-528-7411 or
602-470-1414 for assistance

Moisture never stops
increasing

Other volatiles

Try lower temperature,

see pages 29 & 66

Balance drifting

Call AZI at 800-528-7411 or
602-470-1414 for assistance

No graphical display

Graphics not selected

Press [GRAPH] key at Main
Test Screen

No heat

Voltage selection wrong

Check power module at rear
of instrument, set to correct
voltage. See page 14

Wiring failure in heater circuit

Call AZI at 800-528-7411 or
602-470-1414 for assistance

Heater failure

Call AZI at 800-528-7411 or
602-470-1414 for assistance