4 delta sampling – Campbell Scientific PST3/PST8 Pump and Slug Test System User Manual
Page 23

SECTION 4. TECHNICAL DETAILS
4-2
TABLE 4.2-1. Keller-PSI Calibration Report
Customer: Campbell Scientific
Test Date:
07-06-92
Model No: 169-110-0051
Text Excitation
.5 mA
Serial No: 93907
Test Temperatures:
Room = 27ºC
Pressure Range: 0 to 15 PSIG
Cold = 0ºC
Excitation: .5 mA
Output: 0-60 mV
Test
BFSL
------- Run #1 -------
------- Run #2 -------
------- Run #3 -------
Pressure
Rm Temp
Rm Temp
Error
Rm Temp
Error
Cd Temp
Error
PSIG
Outputs
Outputs
%FSO
Outputs
%FSO
Outputs
%FSO
0.00
0.14
0.176
0.076
0.188
0.101
0.066
-0.006
3.00
9.55
9.519
-0.066
9.522
-0.060
9.449
-0.008
6.00
18.96
18.921
-0.083
18.916
-0.094
18.885
-0.006
9.00
28.37
28.358
-0.025
28.352
-0.038
28.350
-0.002
12.00
37.78
37.801
0.045
37.789
0.020
37.821
0.003
15.00
47.19
47.222
0.069
47.212
0.048
47.265
0.006
12.00
37.78
37.806
0.056
37.796
0.035
37.826
0.004
9.00
28.37
28.361
-0.019
28.356
-0.030
28.352
-0.001
6.00
18.96
18.926
-0.072
18.925
-0.074
18.888
-0.006
3.00
9.55
9.528
-0.047
9.528
-0.047
9.454
-0.008
0.00
0.14
0.189
0.103
0.189
0.103
0.073
-0.005
Maximum Static Error: 0.103 %FSO
Maximum Thermal Error @ Cold:
-.008 %FSO/C
Maximum Non-Repeatability: -0.26 %FSO
Electrical Termination: Black
+Input
Red
+Output
Green
-Output
White
-Input
With 3 wells or more on one PST, the first step
interval will be greater than 0.25 seconds (see
Section 4.4). The elapsed time will be offset by
the increased time required to execute the first
step interval. A change of the execution interval
to 0.5 seconds would result in the following:
Points per
Elapsed Time
Interval
Interval
0 - 40 sec.
0.5 sec.
80
40 - 240 sec.
10 sec.
20
4 - 20 min.
1 min.
16
20 - 200 min.
4 min.
45
200 - 2000 min.
20 min.
90
2000 min. plus
120 min.
2 times the value in *4 location 13 (PST3) or 28
(PST8)
If the interval is changed, it must be a multiple
of 0.125 (1/8) seconds. If the execution interval
were changed to 1 second, the following rate of
data collection would result:
Points per
Elapsed Time
Interval
Interval
0 - 80 sec.
1 sec.
80
80 - 480 sec.
20 sec.
20
8 - 40 min.
2 min.
16
40 - 400 min.
8 min.
45
400 - 4000 min.
40 min.
90
4000 min. plus
240 min.
NOTE: Never set the base interval greater
than 1 second.
4 times the value in *4 location 13 (PST3) or 28
(PST8)
4.4 DELTA SAMPLING
Time and water level are recorded only when
the water level changes by a user-specified
Delta threshold. When using Delta sampling,
each well in the test must have its own Delta
threshold value set in the *4 Table. The default