28 . . . . . . . . . . . Final logarithmic interval in seconds
29 . . . . . . . . . . . Exc. code, transducer #1
30 . . . . . . . . . . . Exc. code, transducer #2
31 . . . . . . . . . . . Exc. code, transducer #3
32 . . . . . . . . . . . Exc. code, transducer #4
33 . . . . . . . . . . . Exc. code, transducer #5
34 . . . . . . . . . . . Exc. code, transducer #6
35 . . . . . . . . . . . Exc. code, transducer #7
36 . . . . . . . . . . . Exc. code, transducer #8
Note: For the final logarithmic interval (13,28) to be in seconds, the loga-
rithmic base interval (3) must be the default 0.25 seconds.
Flip the scan switch on and note the water level reading in each well.
Also check the battery voltage. See Viewing the measurements below.
If the battery voltage is below 11.0 volts (at the start of the test) use the
charger cable to provide power to the battery.
Scanning or during a test, the measurements can be viewed as they are
made.
NOTE: Readings are current ONLY if the scan or record switch is ON.
Key in: *6 [Location #] A
Other keys:
A - advance
B - backup
Scan the
sensors
Viewing the
measurements
Two methods:
Using the CR10
SECTION 3. TEST PROCEDURE
3-5
Excitation Code
0 for sensors from 5-20 psi
1 for sensors over 20 psi