Cr10(x) example program without wind speed, Correction – Campbell Scientific NR-LITE2 Net Radiometer User Manual
Page 23

NR-LITE2 Net Radiometer
 VoltDiff(NR_Wm2,1,mv25,2,True,0,_60Hz,100,0) 
 If WS_ms>=5 Then 
 CNR_Wm2=NR_Wm2*(1+0.021286*(WS_ms-5)) 
 Else 
 CNR_Wm2=NR_Wm2 
 EndIf 
 'Call Data Tables and Store Data 
 CallTable(Table1) 
 CallTable(Table2) 
 NextScan 
EndProg 
7.2.3.3 CR10(X) Example Program without Wind Speed Correction
Shown below is an example program written for the CR10(X) datalogger. 
TABLE 7-5 provides the wiring for the example program. In this example, the 
datalogger measures the output from the sensor every 60 seconds and outputs 
the average net radiation every hour. 
The calibration factor used will only apply for one specific sensor. A new 
value will need to be calculated for every different sensor based on the 
Certificate of Calibration for that sensor (Section 7.2.2, Calibration Factor). 
TABLE 7-5. Wiring for CR10X Example
(without Wind Correction)
Color Function CR10X
Channels
Red Radiation
Signal
1H
Blue Signal
Reference
1L
Jumpered to
AG
Black Shield
G
;{CR10X}
;
*Table 1 Program
01: 60
Execution
Interval
(seconds)
1: Volt (DIFF) (P2)
1: 1
Reps
2: 3**
25 mV Slow Range
;range code for CR10(X) datalogger
3: 1*
DIFF
Channel
4: 1*
Loc [ Net_rad ]
5: 63.29
Mult
;Multiplier for specific sensor
6: 0.0
Offset
;in units of (W m
-2
)/mV (see above)
2: If time is (P92)
1: 0
Minutes (Seconds --) into a
2: 60
Interval (same units as above)
3: 10
Set Output Flag High (Flag 0)
3: Real Time (P77)
1: 110
Day,Hour/Minute
(midnight
=
0000)
4: Average (P71)
1: 1
Reps
2: 1
Loc [ Net_rad ]
17