Cr10(x) example program without wind speed, Correction – Campbell Scientific NR-LITE2 Net Radiometer User Manual

Page 23

background image

NR-LITE2 Net Radiometer

VoltDiff(NR_Wm2,1,mv25,2,True,0,_60Hz,100,0)
If WS_ms>=5 Then

CNR_Wm2=NR_Wm2*(1+0.021286*(WS_ms-5))
Else
CNR_Wm2=NR_Wm2

EndIf
'Call Data Tables and Store Data
CallTable(Table1)

CallTable(Table2)
NextScan
EndProg

7.2.3.3 CR10(X) Example Program without Wind Speed Correction

Shown below is an example program written for the CR10(X) datalogger.
TABLE 7-5 provides the wiring for the example program. In this example, the
datalogger measures the output from the sensor every 60 seconds and outputs
the average net radiation every hour.

The calibration factor used will only apply for one specific sensor. A new
value will need to be calculated for every different sensor based on the
Certificate of Calibration for that sensor (Section 7.2.2, Calibration Factor).

TABLE 7-5. Wiring for CR10X Example

(without Wind Correction)

Color Function CR10X

Channels

Red Radiation

Signal

1H

Blue Signal

Reference

1L

Jumpered to

AG

Black Shield

G

;{CR10X}

;

*Table 1 Program

01: 60

Execution

Interval

(seconds)

1: Volt (DIFF) (P2)

1: 1

Reps

2: 3**

25 mV Slow Range

;range code for CR10(X) datalogger

3: 1*

DIFF

Channel

4: 1*

Loc [ Net_rad ]

5: 63.29

Mult

;Multiplier for specific sensor

6: 0.0

Offset

;in units of (W m

-2

)/mV (see above)

2: If time is (P92)

1: 0

Minutes (Seconds --) into a

2: 60

Interval (same units as above)

3: 10

Set Output Flag High (Flag 0)

3: Real Time (P77)

1: 110

Day,Hour/Minute

(midnight

=

0000)

4: Average (P71)

1: 1

Reps

2: 1

Loc [ Net_rad ]

17