2 datalogger programming, 1 input range and integration, 2 calibration factor – Campbell Scientific NR-LITE2 Net Radiometer User Manual
Page 20: Datalogger programming, Input range and integration, Calibration factor

NR-LITE2 Net Radiometer
TABLE 7-2. Datalogger Connections for Single-Ended Measurement
 
 
 
 
 
Function 
 
 
 
 
 
Color 
 
 
 
CR10X, 
CR510, 
CR500 
 
 
 
 
CR23X, 
21X, CR7 
CR800, 
CR850, 
CR1000, 
CR3000, 
CR5000, 
CR9000(X) 
Radiation Signal Red
S.E. Channel S.E. Channel S.E. Channel
Signal Reference Blue
AG
Shield Black
G
7.2 Datalogger Programming
This section is for users who write their own datalogger 
programs. A datalogger program to measure this sensor can be 
generated using Campbell Scientific’s Short Cut Program 
Builder Software. You only need to read the calibration portion 
of this section if using Short Cut. 
NOTE
The NR-LITE2 outputs a low-level voltage ranging from 0 to a maximum of 
about 15 mV. A differential voltage measurement (VoltDiff() in CRBasic or 
instruction Volt (Diff) (P2) in Edlog) is recommended because it has better 
noise rejection than a single-ended measurement. If a differential channel is 
not available, a single-ended measurement (VoltSE() in CRBasic or instruction 
Volt (SE) (P1) in Edlog) can be used. The acceptability of a single-ended 
measurement can be determined by simply comparing the results of single-
ended and differential measurements made under the same conditions. 
7.2.1 Input Range and Integration
Normally the 15 mV range for the 21X or CR7, the 25 mV range for the 
CR800, CR850, CR1000, CR10(X), or CR500/CR510, and the 50 mV range 
for the CR3000, CR5000, CR9000X, or CR23X are suitable. 
The slow or 60 Hz rejection integration gives a more noise-free reading. A fast 
integration takes less power and allows for faster throughput. 
7.2.2 Calibration Factor
Each NR-LITE2 is provided with a Certificate of Calibration by the 
manufacturer which shows the sensor serial number and a sensitivity or 
calibration factor. This calibration factor, after conversion, is used in the 
datalogger program. Always cross check to ensure that the serial number of 
your NR-LITE2 and the number on the calibration certificate are identical. 
The serial number and sensitivity are also shown on a small label attached to 
the connecting arm near the head of the sensor itself. 
The calibration factor included on the Certificate of Calibration is in units of 
µV/(W m
-2
), which needs to be converted to units of (W m
-2
)/mV for the
multiplier parameter in the datalogger program. To convert the units, divide
14
