Campbell Scientific CS431 Submersible Pressure Transducer User Manual
Page 16

CS431 Submersible Pressure Transducer
5.2.1.3 Example CR10X P
ROGRAM
– High Precision Measurement for 5 PSI Sensors
The Standard and Enhanced Measurement Programming Examples (see
Section 5.2.1.1) result in a resolution of ±0.2 inches. These methods use the 25
mV Slow Range of the CR10X. This level of resolution is acceptable for many
measurement needs. However, in the 0 – 5 PSI range, greater resolution is
often needed.
Using the High Precision Measurement Programming Example (below) will
result in a resolution of ±0.03 inches. This uses the 7.5 mV Slow Range of the
CR10X.
In the example below:
Vr = differential voltage at diff channel 1
Vo = differential voltage at diff channel 2
L = pressure measurement in nominal units
P = pressure measurement in psi units using provided calibration values m
and b from calibration data sheet for the specific sensor being used
T = temperature measurement in degrees C
1: Ex-Del-Diff (P8)
1: 1
Reps
2: 3
25 mV Slow Range
3: 1
DIFF
Channel
4: 1
Excite
all
reps
w/Exchan
1
5: 1
Delay
(units
0.01
sec)
6: 800
mV
Excitation
7: 1
Loc
[
Vr
]
8: 1.0
Mult
9: 0.0
Offset
2: Ex-Del-Diff (P8)
1: 1
Reps
2: 2
7.5
mV
Slow
Range
3: 2
DIFF
Channel
4: 1
Excite
all
reps
w/Exchan
1
5: 1
Delay
(units
0.01
sec)
6: 800
mV
Excitation
7: 2
Loc
[
Vo
]
8: 1.0
Mult
9: 0.0
Offset
; Calculate L factor
L=100*(Vo/Vr)
; Measure Temperature
3: Temp (107) (P11)
1: 1
Reps
2: 5
SE
Channel
3: 2
Excite
all
reps
w/E2
4: 4
Loc
[
T
]
5: 1.0
Mult
6: 0.0
Offset
(cont.)
12