beautypg.com

Campbell Scientific CS431 Submersible Pressure Transducer User Manual

Page 16

background image

CS431 Submersible Pressure Transducer

5.2.1.3 Example CR10X P

ROGRAM

– High Precision Measurement for 5 PSI Sensors

The Standard and Enhanced Measurement Programming Examples (see
Section 5.2.1.1) result in a resolution of ±0.2 inches. These methods use the 25
mV Slow Range of the CR10X. This level of resolution is acceptable for many
measurement needs. However, in the 0 – 5 PSI range, greater resolution is
often needed.

Using the High Precision Measurement Programming Example (below) will
result in a resolution of ±0.03 inches. This uses the 7.5 mV Slow Range of the
CR10X.

In the example below:

Vr = differential voltage at diff channel 1
Vo = differential voltage at diff channel 2
L = pressure measurement in nominal units
P = pressure measurement in psi units using provided calibration values m

and b from calibration data sheet for the specific sensor being used

T = temperature measurement in degrees C

1: Ex-Del-Diff (P8)
1: 1

Reps

2: 3

25 mV Slow Range

3: 1

DIFF

Channel

4: 1

Excite

all

reps

w/Exchan

1

5: 1

Delay

(units

0.01

sec)

6: 800

mV

Excitation

7: 1

Loc

[

Vr

]

8: 1.0

Mult

9: 0.0

Offset


2: Ex-Del-Diff (P8)
1: 1

Reps

2: 2

7.5

mV

Slow

Range

3: 2

DIFF

Channel

4: 1

Excite

all

reps

w/Exchan

1

5: 1

Delay

(units

0.01

sec)

6: 800

mV

Excitation

7: 2

Loc

[

Vo

]

8: 1.0

Mult

9: 0.0

Offset


; Calculate L factor

L=100*(Vo/Vr)

; Measure Temperature

3: Temp (107) (P11)
1: 1

Reps

2: 5

SE

Channel

3: 2

Excite

all

reps

w/E2

4: 4

Loc

[

T

]

5: 1.0

Mult

6: 0.0

Offset

(cont.)

12