beautypg.com

Campbell Scientific CNR2 Radiometer User Manual

Page 14

background image

CNR2 Net Radiometer


; Measure Net Longwave Radiation
2: Volt (Diff) (P2)
1: 1

Reps

2: 22

7.5 mV 60 Hz Rejection Range

3: 2

DIFF

Channel

4: 2

Loc [ Net_LW ]

5: 86.21

Multiplier

6: 0.0

Offset


;
;Measure Net Radiation
3: Z=X+Y (P33)
1: 1

X Loc [ Net_SW ]

2: 2

Y Loc [ Net_LW ]

3: 3

Z Loc [ Net_Rad ]


;Output data to the final storage every 60 minutes
;
4: If time is (P92)
1: 0

Minutes (Seconds --) into a

2: 60

Interval (same units as above)

3: 10

Set Output Flag High (Flag 0)


5: Set Active Storage Area (P80)^2826
1: 1

Final

Storage

Area

1

2: 102

Array

ID


6: Real Time (P77)^11314
1: 1220

Year,Day,Hour/Minute

(midnight

=

2400)


7: Minimum (P74)^13722
1: 1

Reps

2: 10

Value

with

Hr-Min

3: 1

Loc [ Net_SW ]


8: Maximum (P73)^10217
1: 1

Reps

2: 10

Value

with

Hr-Min

3: 1

Loc [ Net_SW ]


9: Average (P71)^10475
1: 1

Reps

2: 1

Loc [ Net_SW ]


10: Minimum (P74)^11228
1: 1

Reps

2: 10

Value

with

Hr-Min

3: 2

Loc [ Net_LW ]


11: Maximum (P73)^17442
1: 1

Reps

2: 10

Value

with

Hr-Min

3: 2

Loc [ Net_LW ]

10