beautypg.com

2 maintenance, 3 calibration, Attributions and references – Campbell Scientific 108-L Temperature Probe User Manual

Page 20: Maintenance, Calibration

background image

Model 108 Temperature Probe

9.2 Maintenance

The 108 probe requires minimal maintenance. For air temperature

measurements, check the radiation shield monthly to make sure the radiation

shield is clean and free from debris. Periodically check cabling for signs of

damage and possible moisture intrusion.

9.3 Calibration

Calibration of the 108 probe is not necessary unless the application requires

removal of the thermistor interchangeability offset described in Section 6,

Specifications. If performing the one point calibration with an Edlog

datalogger, be aware of this precaution:

The value of the offset must be chosen so that the probe outputs the

temperature calculated by the polynomial, not the actual calibration

temperature. For example, a 108 probe placed in a calibration chamber at 0 °C

outputs 0.1 °C. An Offset argument of –0.15

is required for Edlog dataloggers

because at 0 °C, the polynomial calculates a temperature of –0.05 °C

(Appendix C, Conversion of Thermistor Resistance or Voltage Ratio to

Temperature).

10. Attributions and References

Santoprene

®

is a registered trademark of Exxon Mobile Corporation.

EPA installation standard: Quality Assurance Handbook for Air Pollution

Measurement Systems – Volume IV: Meteorological Measurements Version 2.0

WMO standard: WMO No. 8, Seventh edition, 6 Aug 2008 Guide to

Meteorological Instruments and Methods of Observation

14