Model 108 Temperature Probe

outlined in Section 8.3, Electrically Noisy Environments, and Section 8.4, Long

Cable Lengths.

Details of 108 probe measurement and linearization of the thermistor output

are provided in Section 8.2, Measurement and Output Linearization.

7.2.1 CRBasic

The

Therm108() measurement instruction programs CRBasic dataloggers

(CR800-series, CR1000, CR3000, CR5000) to measure the 108 probe. It

supplies a 2500 mV excitation, makes a half-bridge measurement, and converts

the result to temperature using the Steinhart-Hart equation (see Section 8.2,

Measurement and Output Linearization, for more information):

Therm108

(Dest,Reps,SEChan,VxChan,SettlingTime,Integ,Mult,Offset)

Variations:

Temperature reported as °C — set

Mult to 1 and Offset to 0

Temperature reported as °F — set

Mult to 1.8 and Offset to 32

Ac mains noise filtering — set

Integ to _60Hz or _50Hz (see Section

8.3, Electrically Noisy Environments)

Compensate for long cable lengths — Set

SettlingTime to 20000 (see

Section 8.4, Long Cable Lengths)

7.2.2 Edlog

The

AC Half Bridge (P5) measurement instruction programs Edlog

dataloggers (CR10(X), CR510, CR500, CR23X, 21X, and CR7(X)) to measure

the 108 probe. It makes a half-bridge measurement, and the

Polynomial (P55)

instruction converts the result to temperature using a fifth-order polynomial

(see Section 8.2, Measurement and Output Linearization, for more

information):

1: AC Half Bridge (P5)

1: 1

Reps

2: 23

25 mV 60 Hz Rejection Range

3: 1

SE Channel

4: 1

Excite all reps w/Exchan 1

5: 1000

mV Excitation

6: 1

Loc [ T108_C ]

7: 200

Multiplier

8: 0

Offset


2: Polynomial (P55)

1: 1

Reps

2: 1

X Loc [ T108_C ]

3: 1

F(X) Loc [ T108_C ]

4: -26.97 C0

5: 69.635 C1

6: -40.66 C2

7: 16.573 C3

8: -3.455 C4

9: 0.301

C5