8. Troubleshooting and Maintenance
8.1 Troubleshooting
Consult Table 8-1 for recommended actions to resolve various fault conditions.
TABLE 8-1. 020C Wind Direction Sensor Troubleshooting
Symptom Probable
Cause
Solution
Refer
to
No wind
direction
output
Loss of supply
voltage
Check datalogger
12Vdc supply and
connecting cables
Table 7-1, 020C
Wiring
No wind
direction
output
Faulty integrated
circuit amplifier
(output often will
be steady 5.0V)
Replace IC
amplifier (Plug-
in)
Contact
Campbell
Scientific for
repair
authorization
No wind
direction
output
Faulty
potentiometer
Replace
potentiometer
Contact
Campbell
Scientific for
repair
authorization
No change in
wind direction
output below
2-5 mph
Bad bearings
Replace bearings
Contact
Campbell
Scientific for
repair
authorization
8.2 Maintenance Schedule
6-12 Month Intervals:
1. Inspect sensor for proper operation per Section 7.6, Sensor Check-Out.
2. Return to Campbell Scientific for calibration.
12-24 Month Intervals:
1. Return to Campbell Scientific for replacement of sensor bearings.
2. Return to Campbell Scientific for replacement of sensor potentiometer.