Output a reading once, Set output interval – Vaisala HMW90 User Manual
Page 54

User's Guide _______________________________________________________________________
52 ___________________________________________________________________ M211399EN-F
Output a Reading Once
Use the SEND command to output a single measurement message.
SEND
Example:
>send
RH = 21.72 %RH T = 23.12 'C
Set Output Interval
Use the INTV command to change the output interval of the
automatically repeating measurement messages. The measurement
messages are repeated in the RUN mode, or after the R command has
been given.
INTV [n xxx]
where
n
= time interval, range 0 ... 9999.
xxx = time unit = "S", "MIN", or "H"
The shortest output interval (with n = 0) outputs the measurement
messages as quickly as the transmitter produces them, without additional
delay.
Example:
>intv 1 min
Output interval : 1 min