Figure 8, Dewpoint temperature, Table 9 – Vaisala HMM211 User Manual
Page 38: Analog output specifications, Table 10, General specifications

USER'S GUIDE____________________________________________________________________
Figure 8
Dewpoint Temperature
Table 9
Analog Output Specifications
Property Value
Two analog outputs selectable
0 ... 1 V, 0 ... 5 V, 0 ... 10 V
0 ... 20 mA (4 ... 20 mA)
Typical accuracy
±0.1 % full scale
Resolution CH1:
0.025
%RH
CH2: 0.01 % full scale
Typical temperature dependence
0.004 % full scale/°C
Table 10
General Specifications
Property Value
Operating temperature range
Probe
Electronics
-70 ... +180 °C
-5 … +55 °C
Storage temperature range
(electronics)
-40 … +70 °C
Sensor protection:
Standard
Stainless steel sintered filter
Connections
Screw terminals for 0.5 ... 1.5 mm2 wires
38 __________________________________________________________________ M210718EN-B