Example: read rh measurement result, Table 15 get_parameter invoke message example, Table 16 get_parameter response message example – Vaisala HMM105 User Manual
Page 15

Chapter 2 _______________________________________________________________ I2C Interface
VAISALA ____________________________________________________________________ 13
Example: Read RH Measurement Result
Table 15
Get_Parameter Invoke Message Example
Message Bytes (hex) Content
2F
I
2
C address
81
Get_Parameter command
2F
Device address
06
Frame length
4F
RH Parameter ID
6A
CRC high
D4
CRC lo
Table 16
Get_Parameter Response Message Example
Message Bytes (hex) Content
2F
I
2
C address
00
Status byte (no errors, ACK)
81
Get_Parameter command
09
Device address
0B
Frame length
4F
RH Parameter ID
D4
RH value 14.43086624
%RH
E4
66
41
85
CRC high
6A
CRC lo
See also other documents in the category Vaisala Humidifiers:
- Calibration of Digital Transmitters with HMI41 (36 pages)
- Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 (14 pages)
- Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 (30 pages)
- HM34 (30 pages)
- HM40 (47 pages)
- HM44 (52 pages)
- HM70 (83 pages)
- HMD40 (1 page)
- HMD60 (4 pages)
- HMD70 (18 pages)
- HMDW110 (62 pages)
- HMDW80 (51 pages)
- HMI41 (74 pages)
- HMP41 (72 pages)
- HMK15 (39 pages)
- HMM100 (71 pages)
- HMM211 (42 pages)
- HMM212 (36 pages)
- HMM213 (52 pages)
- HMP140 (28 pages)
- HMP155 (84 pages)
- HMP228 (115 pages)
- HMP230 (163 pages)
- HMP240 (130 pages)
- HMP260 (118 pages)
- HMP60 (71 pages)
- HMT100 (52 pages)
- HMT120 (87 pages)
- HMT130 (95 pages)
- HMT140 (76 pages)
- HMT310 (88 pages)
- HMT310 (105 pages)
- HMT330 (209 pages)
- HMT360 (97 pages)
- HMT360 (63 pages)
- HMT360N (110 pages)
- HMW40 (1 page)
- HMW90 (110 pages)
- SHM40 (68 pages)
- RDP100 (14 pages)