Maintenance, 3 calibration table – Vaisala HMD70 User Manual

Page 13

background image

HMD70U/Y Transmitters

U303en-1.1

Operating Manual

1997-07-31

9

4.3

Calibration table

Temperature

°C

15

20

25

30

35

°F

59

68

77

86

95

LiCl

%RH

*)

11.3

11.3

11.3

11.3

4...20 mA

mA

5.81

5.81

5.81

5.81

0...20 mA

mA

2.26

2.26

2.26

2.26

0...1 V

V

0.113

0.113

0.113

0.113

0...5 V

V

0.565

0.565

0.565

0.565

0...10 V

V

1.13

1.13

1.13

1.13

NaCl

%RH

75.6

75.5

75.3

75.1

74.9

4...20 mA

mA

16.10

16.08

16.05

16.02

15.98

0...20 mA

mA

15.12

15.10

15.06

15.02

14.98

0...1 V

V

0.756

0.755

0.753

0.751

0.749

0...5 V

V

3.780

3.775

3.765

3.755

3.745

0...10 V

V

7.56

7.55

7.53

7.51

7.49

*)

LiCl solution must not be used or stored in temperatures below +18

°

C (+64

°

F);

otherwise the equilibrium humidity of the salt solution changes permanently

Table 1

Greenspan’s calibration table with output values according to
the chosen scale

5.

MAINTENANCE

5.1

Replacing the HUMICAP

180 sensor and the filter

Remove the damaged sensor and insert a new one. Handle the sensor by the
plastic socket. DO NOT TOUCH THE SENSOR PLATE. Recalibrate the
transmitter.

Replace a dirty filter (membrane or sintered) to ensure a maximum lifetime
and a fast response for the sensor. Do not try to clean the filter.