Technical Data

Performance

Relative Humidity
Measurement range

0 ... 100 %RH

Accuracy (including non-linearity,
hysteresis, and repeatability)

at 0 ... +40 °C (+32 ... +104 °F)

±1.5 %RH (0 ... 90 %RH)

±2.5 %RH (90 ... 100 %RH)

at -10 ... 0 °C and +40 ... +60 °C
(+14 ... +32 °F and +104 ... +140 °F)

±3.0 %RH (0 ... 90 %RH)

±4.0 %RH (90 ... 100 %RH)

Factory calibration uncertainty
at +20 °C (+68 °F)

±1.5 %RH

Humidity measurement response time
(90%) with plastic grid filter

17 s

Stability

±2 %RH over 2 years

Humidity sensor

HUMICAP® 180R

Temperature
Measurement range

-10 ... +60 °C (+14 ...+140 °F)

Accuracy over temperature range:

at 0 ... +40 °C (+32 ... +104°F)

±0.2 °C (0.36 °F)

at -10 ... 0 °C and +40 ... +60 °C
(+14 ... 32 °F and +104 ... +140 °F) ±0.4 °C (0.72 °F)

Temperature Sensor

Pt1000 RTD Class F0.1
IEC 60751