Vaisala Calibration of Digital Transmitters with HMI41 User Manual
Page 34

User's Guide ______________________________________________________________________
32 __________________________________________________________________ M210777EN-B
RH
Insert the transmitter probe to the lower
humidity.
RH
%
Let the readings stabilize.
MODE
RH
set
%
Set the blinking RH reading to the correct
value (e.g. to the equilibrium RH of the
salt solution).
V (number up) or
W (number down)
ENTER
RH
Insert the transmitter probe to the higher
humidity.
RH
%
Let the readings stabilize.
MODE
RH
%
set
Set the blinking RH reading to the correct
value (e.g. to the equilibrium RH of the
salt solution).
▲ (number up) or
▼ (number down)
ENTER to conclude
RH
Calibration has been successful.
ON/OFF
(disconnect the cable)
RH
Calibration has not been successful.
Perform the calibration again.