Two-point calibration with hmi41 as a terminal – Vaisala Calibration of Digital Transmitters with HMI41 User Manual

Page 33

background image

Appendix A _______________________________________________Quick Reference to Calibration

VAISALA _______________________________________________________________________ 31

RH

Insert the HMI41 probe and the
transmitter probe to the higher humidity.

RH

%

Let the readings stabilize. Conclude the
calibration. Alternatively, you can change
the display to show the difference in
readings.

ENTER to conclude or
HOLD to have the
difference on the display

RH

%

Let the readings stabilize. Conclude the
calibration or return to the previous
display.

ENTER to conclude or
HOLD to return to the
previous display

RH

Calibration has been successful.

ON/OFF
(disconnect the cable)

RH

Calibration has not been successful.
Perform the calibration again.

Two-Point Calibration with HMI41 as a Terminal

DISPLAY DESCRIPTION

PRESS

RH

set min max

hyst hold

Lo bat Hi

Pabs

Pa

g/m

T

Td

°C°F

°C°F

%

HMI41 has been turned on.

Indication of the software version (if the
version is 1.02 or more).

bat Hi

Indication of the battery voltage.

RH

Select the desired calibration mode.

▲ (number up) or
▼ (number down)
ENTER