* With the " REC. " symbol on the display :

a) Press the " REC Button " ( 3-8, Fig. 1 ) once, the "

REC. MAX. " symbol along with the maximum value
will appear on the display.
If intend to delete the maximum value, just press
the " Hold Button " ( 3-4, Fig. 1 ) once, then the
display will show the " REC. " symbol only & execute
the memory function continuously.

b)Press the " REC Button " ( 3-8, Fig. 1 ) again, the

" REC. MIN. " symbol along with the minimum value
will appear on the display.
If intend to delete the minimum value, just press
the " Hold Button " ( 3-4, Fig. 1 ) once, then
the display will show the " REC. " symbol only &
execute the memory function continuously.

5. CALIBRATION PROCEDURE

1)The meter can be calibrated under two kinds standard

solution :

0 NTU standard solution
100 NTU standard solution

2)The meter ship along with

a. Testing bottle with 0 NTU standard solution x 1 PC
b. Testing bottle with 100 NTU standard solution x 1 PC

as the standard accessories.

3)The complete calibration should be executed by

following two solution :

0 NTU standard solution
100 NTU standard solution

7