Then the Display will show :

100

CAL

ntu

Now the meter is finished the " 0 NTU " calibration
procedures and ready for " 100 NTU " calibration
procedures.

7) 100 NTU calibration

When finish the " 0 NTU " calibration procedures and
the Display show :

100

CAL

ntu

The meter is ready for " 100 NTU " calibration.

* Insert the " 100 NTU standard solution into the

" Container " ( 3-2 , Fig. 1) and envelope in the " Cover "
( 3-1, Fig. 1 ) completely.
Press " CAL Button " ( 3-5, Fig. 1 ) once, the
Display will show following text with flashing ( approx.
10 seconds ).

flashing

CAL

100

ntu