Sper Scientific LUTU-2016 Turbidity Meter User Manual

Page 11

background image

b. Press " CAL Button " ( 3-5, Fig. 1 ) continuously

(at least 3 seconds ) until the Display show the text
" CAL " then release the button.

CAL

Fig. 5-1

ntu

c. After the LCD show above Fig. 5-1, wait a while

the display will show

0.00

CAL

ntu

Now the meter is ready for the " 0 NTU " calibration

d. Press " CAL Button " ( 3-5, Fig. 1 ) once, the

Display will show following text with flashing ( approx.
10 seconds ).

flashing

CAL

0.0

ntu

9