8

the object with the IR probe, adjust the
“Emissivity Adjust VR” (Fig.1-10) with a small
flat head screwdriver until the readings
match.

Turning the VR to the left (counter-
clockwise) will adjust the emissivity to
0.10 to 0.94.

Turn the VR all the way to the right to
reset the probe to 0.95.

FACTORS AFFECTING

MEASUREMENT

1. Measurement Field/Distance:

The ideal distance between the probe
and the surface being measured is
10 mm (3/8 inch)

The size of the object should be at
least 2.5 mm (3/32 inch)

2. Positioning: For the most accurate

reading, the probe should be held as
perpendicular to the object as possible.

5

MEASUREMENT PROCEDURE

1. Plug the IR probe plug (Fig.1-9) into the

meter input socket (Fig.1-13).


2. Turn meter on using the POWER switch

(Fig.1-6).


3. Choose scale using the °C/°F button

(Fig.1-3). The chosen scale will appear
in the lower right of the display.

4. Select the resolution by moving the

1°/0.1° switch (Fig.1-7). Use “1°” for
general use.

5. While holding the IR probe main body

(Fig.1-11), point the tip of the probe
(Fig.1-12) at the object being measured.
The meter will display the surface
temperature.

The ideal distance between the probe
and the surface being measured is
10 mm (3/8 inch)

The size of the object should be at
least 2.5 mm (3/32 inch)