EA001A THERMOCOUPLE SIMULATION & MEASUREMENT ADAPTER
OPERATION MANUAL
Transmille Ltd.
Page 15
Automatic Cold Junction Compensation
Click on the softkey marked
This activates the built in temperature sensor in the thermocouple plug to accurately
compensate for the temperature at the point of connection.
Manual Cold Junction Compensation
Click on the softkey marked
Use the calibrator keypad to key in the required CJC temperature, for example 24.5°C :
Manual cold junction mode is ideal for checking the temperature sensor in
the temperature meter (UUT) to ensure it is correctly measuring the cold
junction temperature. Simply set the manual cold junction temperature to
zero in order to allow the temperature meter to indicate the simulated temperature
+ CJC temperature (as measured by the UUT).