beautypg.com

頁面 18 – Dwyer 8912 User Manual

Page 18

background image

17

Intercept (value Z)
1. While press key save the S
setting, the meter will enter intercept
input automatically. The last digit
will flash to indicate the Z value is
now ready for edit.
2. If the last saved Z value is different
from your paper, press key to
move to the digit which you want
to edit and then use key to
change the value.
3. Press key to save new input
Z value and then meter will be back
to normal RH value mode.

Important Note: While setting Z value,
only input 3 digits after decimal point, so
user need to round off the 4th digit.

Important Note: While setting the slope,
only input 5 digits after decimal point, so
user need to round off the 6th digit.
User should also input two digits before
decimal.

Example:
a. If the S value is 8.239678, user
should input 08 23968
b. If the S value is 11.23968, user
should input 11 23968
c. If the S value is 0.2396838, user
should input 00 23968 (Fig. S)

(Fig. S)