Blue-white, Temperature vs. pressure pressure temperature and – Blue-White F-430N Machined Acrylic Tube User Manual
Page 3
![background image](/manuals/371366/3/background.png)
Temperature vs. Pressure
Pressure
Temperature
and
PSIg / BAR
Temperature
130°F / 54.4°C
120°F / 48.9°C
110°F / 43.3°C
100°F / 37.8°C
90°F / 32.2°C
140°F / 60°C
80°F / 26.7°C
70°F / 21.1°C
0 / 0
25 / 1.7
50 / 3.5
75 / 5.2
100 / 6.9
130 / 9
Pressure and temperature limits are inversely proportional. At the maximum suggested pressure the tempera-
ture should approach 70°F / 21.1°C; at the maximum suggested temperature the pressure should approach
zero psi. We cannot guarantee our flowmeters will not be damaged either at or below the suggested limits
simply because of many factors which influence meter integrity; stress resulting from meter misalignment,
damage due to excessive vibration and/or deterioration caused by contact with certain chemicals as well as
direct sunlight. These situations and others tend to reduce the strength of the materials from which the meters
are manufactured.
Flowmeters are tested and calibrated for water or air only.
Although meters may be suitable for other chemicals, Blue-White cannot guarantee their suitability. It
is the responsibility of the user to determine the suitability of the flowmeter in their application.
Application Note
Industries, Ltd.
Blue-White
Industries, Ltd.
®