Range specific specifications – Alicat MC Series Mass Flow Controller User Manual
Page 59
59
Technical Data for MCV & MCVS Mass Flow Controllers
0 to 0.5 sccm Full Scale through 0 to 20 slpm Full Scale
The Alicat MCV mass flow controller is designed for applications that require tight shut-off such as vacuum
coating and sputtering processes. An integrated pneumatic shut-off valve is normally closed and provides
positive shut-off of 1 x 10
-9
atm scc/sec Helium max.
MCVS controllers are for use with aggressive gases.
Performance
MCV Mass Flow Controller
MCVS Mass Flow Controller
Accuracy at calibration conditions
after tare
± (0.8% of Reading + 0.2% of Full Scale)
High Accuracy at calibration conditions
after tare
± (0.4% of Reading + 0.2% of Full Scale)
High Accuracy option not available for units ranged under 5 sccm.
Repeatability
± 0.2% Full Scale
Zero Shift and Span Shift
0.02% Full Scale / ºCelsius / Atm
Operating Range / Turndown Ratio
0.5% to 100% Full Scale / 200:1 Turndown 1% to 100% Full Scale / 100:1 Turndown
Maximum Controllable Flow Rate
102.4% Full Scale
Typical Response Time
100 ms (Adjustable)
Warm-up Time
< 1 Second
Integrated Valve Leak Integrity
1 x 10
-9
atm scc/sec Helium max
Standard Specifications (Contact Alicat for available options.)
Operating Conditions
MCV Mass Flow Controller
MCVS Mass Flow Controller
Mass Reference Conditions (STP)
25ºC & 14.696 psia (standard — others available on request)
Operating Temperature
−10 to +50 ºCelsius
Humidity Range (Non–Condensing)
0 to 100%
Maximum Pressure
145 psig
Mounting Attitude Sensitivity
None
Valve Type
Normally Closed
Ingress Protection
IP40
Wetted Materials
MCV: 316L,303 & 302 Stainless Steel, Viton®, Silicone RTV (Rubber), Glass Reinforced
Nylon, Aluminum, Brass, 430FR Stainless Steel, Silicon, Glass, PCTFE.
MCVS: 316LSS, 303SS, 430FRSS, FFKM (Kalrez) standard, Viton, EPDM, Buna, Neo-
prene as needed for some gases.
If your application demands a different material, please contact Alicat.
Communications / Power
Whisper MCV & MCVS Mass Flow Controller
Monochrome LCD or Color TFT
Display with integrated touchpad
Simultaneously displays Mass Flow, Volumetric Flow, Pressure and Temperature
Digital Input/Output Signal
1
Options
RS-232 Serial / RS-485 Serial / PROFIBUS
3
Analog Input/Output Signal
2
Options
0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA
Optional Secondary Analog
Input/Output Signal
2
0-5 Vdc / 1-5 Vdc / 0-10 Vdc / 4-20 mA
Electrical Connection Options
8 Pin Mini-DIN / 9-pin D-sub (DB9) / 15-pin D-sub (DB15) / 6 pin locking
Supply Voltage
12 to 30 Vdc (15-30 Vdc for 4-20 mA outputs)
Supply Current
0.250 Amp
1. The
Digital Output Signal communicates Mass Flow, Volumetric Flow, Pressure and Temperature
2. The
Analog Output Signal and Optional Secondary Analog Output Signal communicate your choice of Mass
Flow, Volumetric Flow, Pressure or Temperature
3. If selecting
PROFIBUS, no analog signal is available. PROFIBUS units do not have the display. See PROFIBUS
specifications for PROFIBUS supply voltages and currents. (
www.alicat.com/profibus)
Range Specific Specifications
Full Scale Mass Flow Controller
Mechanical Dimensions
Process Connections
MCV 0.5SCCM to 20SLPM
4.8”H x 6.8”W x 1.5”D
1/4” VCR® Male
MCVS 0.5SCCM to 20SLPM
5.5”H x 6.8”W x 1.5”D
1/4” VCR® Male
Welded VCR® process connections are recommended for MCV and MCVS applications. Please contact Alicat.