Alicat V Series Volumetric Flow Meter User Manual

Page 34

background image

32

Dimensional Drawings: page 35, 36

Specification

Mass

Meter

Volumetric

Meter

Description

Accuracy

± (0.8% of Reading + 0.2% of Full Scale)

At calibration conditions after tare

High Accuracy Option

± (0.4% of Reading + 0.2% of Full Scale)

At calibration conditions after tare

Accuracy: Bi-directional

Meters Only

± (0.8% of reading + 0.2% of total span

positive full scale to negative full scale)

At calibration conditions after tare

Repeatability

± 0.2%

Full Scale

Operating Range

1% to 100% Full Scale

Measure

Typical Response Time

10

Milliseconds (Adjustable)

Standard Conditions (STP)

25ºC & 14.696PSIA

Not Applicable

Mass Reference Conditions

Operating Temperature

−10 to +50

ºCelsius

Zero Shift

0.02%

Full Scale / ºCelsius / Atm

Span Shift

0.02%

Full Scale / ºCelsius / Atm

Humidity Range

0 to 100%

Non–Condensing

Measurable Flow Rate

128%

Full Scale

Maximum Pressure

145

145

1

PSIG

Output Signal Digital

Mass, Volume,

Pressure & Temperature

Volumetric Flow

RS-232 Serial

Output Signal Analog

Mass Flow

Volumetric Flow

0-5Vdc

Optional Output Signal

Secondary Analog

Mass, Volumetric

Pressure or Temperature

Volumetric Flow

0-5 Vdc or 0-10Vdc

or 4-20mA

Electrical Connections

8 Pin

Mini-DIN

Supply Voltage

7 to 30 Vdc (15-30Vdc for 4-20mA outputs)

Supply Current

0.035Amp (+ output current on 4-20mA)

Mounting Attitude Sensitivity

0%

Tare after installation

Warm-up Time

< 1

Second

Wetted Materials

2

303 & 302 Stainless Steel, Viton®, Silicone RTV (Rubber), Glass Reinforced

Nylon, Aluminum.

1.

Volumetric meters only: Operating pressure limitations determined by Reynolds number thresholds. For

operating pressures >10PSIG, please contact Application Assistance for more details.

2. If your application demands a different material, please contact Application Assistance for available options.

The following specifications are for the standard configuration of the Alicat product as shipped from the

factory. There are many low-cost customization options available.

Full Scale Flow

Mass Meter

Full Scale Flow

Volumetric Meter

Mechanical

Dimensions

Process

Connections

1

Pressure Drop

2

(PSID)

100SCCM to 10SLPM

100CCM to 10LPM

4.1”H x 2.4”W x 1.1”D

1/8” NPT Female

1.0

20SLPM

20LPM

4.2”H x 2.4”W x 1.1”D

1/8” NPT Female

1.0

Compatible with Beswick®, Swagelok® tube, Parker®, face seal, push connect and compression adapter

1.

fittings.

Venting to atmosphere. Lower Pressure Drops Available, please contact Application Assistance.

2.

Mechanical Specifications

Technical Data for Low Flow Mass & Volumetric Flow Meters

0 to 100 SCCM Full Scale through 0 to 20 SLPM Full Scale