Alicat V Series Volumetric Flow Meter User Manual
Page 19
17
Flashing Error Message: Our volumetric flow meters and controllers display an error message
(VOV = volumetric overrange) when a measured parameter exceeds the range of the sensors in the
device. When any item flashes on the display, the flashing parameter is not accurate. Reducing the
value of the flashing parameter to within specified limits will return the unit to normal operation and
accuracy.
Select Menu Mode
Pushing “Mode” once will bring up the “Select Menu” display (Figure 7, page 12). Push the button
nearest your selection to go to the corresponding screen. Push “Mode” again to return to the Main
Mode display. (
Note: If your meter was ordered with Totalizing Mode option (page 38), pushing the
“Mode” button once will bring up the “Totalizing Mode” display. Pushing “Mode” a second time will bring
up the “Select Menu” display.)
Gas Select Mode
The Gas Select Mode is accessed by pressing the button above “Gas Select” on the Select Menu
display. The screen will appear as shown in Figure 8 (page 12). The selected gas is displayed on the
default main mode screen as shown in Figure 11, and is indicated by the arrow in the gas select mode
screen in Figure 8. To change the selected gas, use the buttons under “UP” and “DOWN” or those
above “PgUP” and “PgDWN” to position the arrow in front of the desired gas. When the mode is cycled
back to the main mode, the selected gas will be displayed on the main screen.
Note: Gas Select Mode may not be available for units ordered for use with a custom gas or blend.
Communication Select Mode
The Communication Select mode is accessed by pressing the button below “Comm. RS-232” on the
Select Menu display. Please see page 13 for Communication Select mode instructions.
Manufacturer Data Mode
“Manufacturuer Data” is accessed by pressing the “Mfg. Data” button on the Select Menu display
(Figure 7,page 12). The “Mfg 1” display shows the name and telephone number of the manufacturer.
The“Mfg 2” display shows important information about your flow meter including the model number,
serial number, and date of manufacture (Figure 10, page 14).
Miscellaneous Mode
The Miscellaneous mode is accessed by pressing the button above the “Misc” label in the upper right
hand corner of the Select Menu display. The screen will appear as shown in Figure 11. Push the
button above “Select” to move the cursor even with the item you wish to adjust. Then use the “UP” and
“DOWN” buttons to make the adjustment. See pages 14 & 15.
NOTE: All Miscellaneous changes are recorded when you exit the Miscellaneous display.
RS-232 Output and Input
Configuring HyperTerminal®:
Open your HyperTerminal® RS-232 terminal program (installed under the “Accessories” menu on
1.
all Microsoft Windows® operating systems).
Select “Properties” from the file menu.
2.
Click on the “Configure” button under the “Connect To” tab. Be sure the program is set for: 19,200
3.
baud (or matches the baud rate selected in the RS-232 communications menu on the meter) and
an 8-N-1-None (8 Data Bits, No Parity, 1 Stop Bit, and no Flow Control) protocol.
Under the “Settings” tab, make sure the Terminal Emulation is set to ANSI or Auto Detect.
4.
Click on the “ASCII Setup” button and be sure the “Send Line Ends with Line Feeds” box
5.
is not
checked and the “Echo Typed Characters Locally” box and the “Append Line Feeds to Incoming
Lines” boxes are checked. Those settings not mentioned here are normally okay in the default
position.
Save the settings, close HyperTerminal® and reopen it.
6.