beautypg.com

Alicat V Series Volumetric Flow Meter User Manual

Page 12

background image

10

M Series Mass Flow Meter Operation

The M Series Mass Flow Meter provides a multitude of useful flow data in one simple, rugged device.

The M Series can have several screen “modes” depending on how the device is ordered. All M Series

meters have a default Main Mode, Select Menu Mode, a Gas Select Mode (the Gas Select Mode

may not be available on meters calibrated for a custom gas or blend), a Communication Select Mode,

Manufacturer Data Mode and a Miscellaneous Mode. In addition, your device may have been ordered

with the optional Totalizing Mode (page 38). The device defaults to Main Mode as soon as power is

applied to the meter.
Main Mode
The main mode screen defaults on power up with the mass flow on the primary display. The following

parameters are displayed in the main mode as shown in Figure 6:

PSIA

o

C Tare

+13.49 +22.73

MASS

SCCM

Air

+0.000 +0.000

Volume Mass Main

Figure 6. Main Mode Display, M Series Flow Meter

The “MODE” button in the lower right hand corner toggles the display between the Main display and the

Select Menu display.
Tare – Pushing the dynamically labeled “Tare” button in the upper right hand corner tares the flow meter

and provides it with a reference point for zero flow. This is a simple but important step in obtaining

accurate measurements. It is good practice to “zero” the flow meter each time it is powered up. If the

flow reading varies significantly from zero after an initial tare, give the unit a minute or so to warm up

and re-zero it.
If possible, zero the unit near the expected operating pressure by positively blocking the flow downstream

of the flow meter prior to pushing the “Tare” button. Zeroing the unit while there is any flow will directly

affect the accuracy by providing a false zero point. If in doubt about whether a zero flow condition exists,

remove the unit from the line and positively block both ports before pressing the “Tare” button. If the unit

reads a significant negative value when removed from the line and blocked, it is a good indication that

it was given a false zero. It is better to zero the unit at atmospheric pressure and a confirmed no flow

conditions than to give it a false zero under line pressure.
Note: A remote tare can be achieved by momentarily grounding pin 4 to tare as shown in Figure 5 on

page 9.