Thermo Fisher Scientific Ion Selective Electrodes Potassium User Manual

Page 16

background image

Instruction Manual

Potassium Electrode

16


Drift (reading

samples and standards

allow solutions to come to room

slowly changing

at different temperatures

temperature before measurement

in one direction)

electrode

exposed

soak

electrode

in

potassium

to

interferences standard


incorrect

reference

use

recommended

filling

solution

filling

solution



Low Slope or

standards contaminated

prepare fresh standards

No Slope

or incorrectly made


ISA

not

used

use

recommended

ISA


standard

used

as

ISA

use

ISA


defective

electrode

check

electrode

operation


electrode

exposed

soak

electrode

standard

to

interferences

in

potassium

air bubble on membrane

remove bubble by redipping probe



Noisy or Unstable

defective meter

check meter with shorting strap

Readings (readings

continuously or

air bubble on membrane

remove bubble by redipping

randomly

changing)

electrode


defective

electrode

replace

electrode

electrode

exposed

soak

electrode

in

potassium

to

interferences standard


meter or stirrer

ground meter or stirrer

not

grounded

outer filling solution

fill electrode to level just

level

too

low

below

fill

hole