Instruction Manual

Potassium Electrode

15

Technique

Be sure that the electrode's limit of detection has not been exceeded. Be sure that the analysis
method is clearly understood and that Good Laboratory Practice has been followed.

Refer to the instruction manual again. Reread sections

GENERAL PREPARATION

and

ELECTRODE

CHARACTERISTICS.


If trouble still persists, call Eutech Instruments Pte Ltd. at (65) 6778-6876 and ask for the Customer
Services Department.



TROUBLESHOOTING HINTS

Symptom

Possible Causes

Next Step

Out of Range

defective meter

check meter with shorting strap

Reading (see

meter

instruction

manual)


electrode(s) not

unplug electrode(s) and reset

plugged

in

properly


reference

electrode

be

sure

reference

electrode

is

not

filled

filled

air

bubbles

on

remove

bubble

by

membrane

redipping

electrode


electrodes

not

put

electrode(s)

in

solution

in solution



"Incorrect Answer"

incorrect scaling

plot millivolts on the linear axis.

(but calibration

of semi-log paper

On the log axis, be sure

curve

is

good)

concentration

numbers

within

each

decade

are

increasing

with

increasing

concentration


incorrect sign

be sure to note sign of millivolt

number correctly


incorrect

standards

prepare

fresh

standards

wrong

units

used

apply

correct

conversion

factor:

10

-3

M = 39 ppm K

+1

= 75 ppm as KCl


sample

carryover

rinse

electrodes

thoroughly

between samples