Thermo Fisher Scientific Ion Selective Electrodes Potassium User Manual
Page 15

Instruction Manual
Potassium Electrode
15
Technique
Be sure that the electrode's limit of detection has not been exceeded. Be sure that the analysis
method is clearly understood and that Good Laboratory Practice has been followed.
Refer to the instruction manual again. Reread sections
GENERAL PREPARATION
and
ELECTRODE
CHARACTERISTICS.
If trouble still persists, call Eutech Instruments Pte Ltd. at (65) 6778-6876 and ask for the Customer
Services Department.
TROUBLESHOOTING HINTS
Symptom
Possible Causes
Next Step
Out of Range
defective meter
check meter with shorting strap
Reading (see
meter
instruction
manual)
electrode(s) not
unplug electrode(s) and reset
plugged
in
properly
reference
electrode
be
sure
reference
electrode
is
not
filled
filled
air
bubbles
on
remove
bubble
by
membrane
redipping
electrode
electrodes
not
put
electrode(s)
in
solution
in solution
"Incorrect Answer"
incorrect scaling
plot millivolts on the linear axis.
(but calibration
of semi-log paper
On the log axis, be sure
curve
is
good)
concentration
numbers
within
each
decade
are
increasing
with
increasing
concentration
incorrect sign
be sure to note sign of millivolt
number correctly
incorrect
standards
prepare
fresh
standards
wrong
units
used
apply
correct
conversion
factor:
10
-3
M = 39 ppm K
+1
= 75 ppm as KCl
sample
carryover
rinse
electrodes
thoroughly
between samples