beautypg.com

Thermo Fisher Scientific Ion Selective Electrodes Nitrate User Manual

Page 18

background image

Instruction Manual

Nitrate Ion Electrode

16

SPECIFICATIONS

Concentration Range:

1M to 7 x 10

-6

M

(6.2 x 10

4

to 0.5 ppm as NO

3

-1

)


pH Range:

2.5 to 11


Temperature Range:

0

o

to 40

o

C


Resistance:

100 Mohms


Reproducibility:

±2%


Samples:

aqueous solutions only

no organic solvents


Size:

110 mm in length; 12 mm in diameter; 1 m cable length


Storage:

Store in dilute nitrate solution



ORDERING INFORMATION

CODE NO.

DESCRIPTION


EC-NO3-03

Nitrate Ion Combination Electrode, epoxy body


EC-SCS-NT1-BT Nitrate Standard, 0.1M NaNO

3


EC-SCS-NT2-BT Nitrate Standard. 1,000 ppm NO

3

-1


EC-SCS-NT3-BT Nitrate Standard. 100 ppm NO

3

-1


EC-ISA-NT1-BT Nitrate ISA (Ionic Strength Adjuster), 2 M (NH

4

)

2

SO

4