Thermo Fisher Scientific Ion Selective Electrodes Nitrate User Manual
Page 16

Instruction Manual
Nitrate Ion Electrode
14
TROUBLESHOOTING HINTS
Symptom
Possible Causes
Next Step
Out of Range
defective meter
check meter with shorting strap
Reading
(see
meter
instruction
manual)
defective electrode
check electrode operation
electrodes not
unplug electrodes and reset
plugged in properly
reference electrode
be sure reference
not filled
electrode is filled
air bubble on
remove bubble by
membrane
re-dipping electrode
electrodes not in
put electrodes in
solution
solution
Noisy or Unstable defective meter
check meter with
Readings (readings
shorting strap
continuously or
rapidly changing) air bubble on
remove bubble by
membrane
re-dipping electrode
defective electrode
replace electrode
ISA not used
use recommended ISA
meter or stirrer
ground meter or stirrer
not grounded
electrode exposed
soak electrode in
to interferences
nitrate standard
outer filling
fill electrode to
solution level
level just below
too low
the fill hole
Drift (reading
samples and standards
allow solutions to come to room
slowly changing at different temperatures
temperature in one direction) before
measurement
electrode exposed
soak electrode in
to interferences
nitrate standard
incorrect reference
use recommended
filling solution
filling solution