Thermo Fisher Scientific Ion Selective Electrodes Nitrate User Manual

Page 16

background image

Instruction Manual

Nitrate Ion Electrode

14

TROUBLESHOOTING HINTS

Symptom

Possible Causes

Next Step

Out of Range

defective meter

check meter with shorting strap

Reading

(see

meter

instruction

manual)


defective electrode

check electrode operation


electrodes not

unplug electrodes and reset

plugged in properly


reference electrode

be sure reference

not filled

electrode is filled


air bubble on

remove bubble by

membrane

re-dipping electrode


electrodes not in

put electrodes in

solution

solution


Noisy or Unstable defective meter

check meter with

Readings (readings

shorting strap

continuously or
rapidly changing) air bubble on

remove bubble by

membrane

re-dipping electrode


defective electrode

replace electrode


ISA not used

use recommended ISA


meter or stirrer

ground meter or stirrer

not grounded


electrode exposed

soak electrode in

to interferences

nitrate standard


outer filling

fill electrode to

solution level

level just below

too low

the fill hole


Drift (reading

samples and standards

allow solutions to come to room

slowly changing at different temperatures

temperature in one direction) before

measurement


electrode exposed

soak electrode in

to interferences

nitrate standard


incorrect reference

use recommended

filling solution

filling solution