Instruction Manual

Nitrate Ion Electrode

15

Low Slope or

standards contaminated

prepare fresh standards

No Slope

or incorrectly made


ISA not used

use recommended ISA


standard used as ISA

use ISA


electrode exposed

soak electrode in

to interferences

nitrate standard


defective electrode

check electrode operation


air bubble on

remove bubble by

membrane

re-dipping probe


"Incorrect Answer" incorrect scaling

plot millivolts on the linear axis. On the

(but calibration

of semi-log paper

log axis be sure concentration numbers

curve

is

good)

within

each

decade

are

increasing

with

increasing

concentration.

.

incorrect sign

be sure to note sign of millivolt number

correctly


incorrect standards

prepare fresh standards


wrong units used

apply correct conversion factor:

10

-3

M

=

62.0

ppm

as

NO

3

= 14.0 ppm as N


sample carryover

rinse electrodes thoroughly between

samples