beautypg.com

Thermo Fisher Scientific Ion Selective Electrodes Ammonia User Manual

Page 22

background image

value correctly


incorrect standards

prepare fresh standards


wrong units used

apply correct conversion factor:

10-

3

M = 17 ppm as

NH

3

= 14 ppm as N


buffer added to

add same proportions of buffer to

standards and not

standards and samples

samples

sample carryover

rinse electrodes thoroughly

between samples


Checking the Electrode Inner Body


If the electrode slope is found to be low during operation, the following solutions will be necessary
to check the inner body:

- pH 4 Buffer (0.1M NaCl added)
Add 2.9 grams of reagent-grade NaCl to 500 ml of pH 4 buffer. Dissolve the solid. The solution
may be stored for repeated use.

- pH 7 Buffer (0.1M NaCl added)
Add 2.9 grams of reagent-grade NaCl to 500 ml of pH 7 buffer. Dissolve the solid. The solution
may be stored for repeated use.

Disassemble the ammonia electrode. If the electrode is dry, soak the glass tip of the inner body in
Eutech Ammonia Electrode Filling Solution (Code No. EC-RE-032) for at least two hours.

Rinse the electrode thoroughly with distilled water. Put 100 ml of pH 7 buffer (0.1M NaCl added)
in a 150 ml beaker. Place the beaker on the magnetic stirrer, and begin stirring. Immerse the tip of
the inner body in the solution so that the reference element is covered. Make sure that the meter is
in the mV mode. Record the meter reading when stable.

Rinse the inner body thoroughly in distilled water. Put 100 ml of pH 4 buffer (0.1M NaCl added) in
a 150 ml beaker, place the beaker on the magnetic stirrer, and begin stirring. Immerse the tip of the
inner body in the solution so that the reference element is covered. Observe the change in the meter
reading carefully. In less than 30 seconds after immersion, the reading should change 100 mV. The
meter reading should stabilize in 3 - 4 minutes, with a difference greater than 150 mV if the inner
body sensing elements are operating properly.