Thermo Fisher Scientific Ion Selective Electrodes Ammonia User Manual

Page 21

background image

high temperatures


membrane failure

replace membrane

(wet, perforation,

discoloration)


samples and standards

allow samples and standards to come

standards not at

to room temperature before use

constant temperature


heat generated by

place insulating material between

magnetic stirrer

stirrer and beaker


defective inner body

refer to

Checking the Electrode

Inner Body


electrode exposed

hold electrode by outer body and

to air for extended period

pull on electrode cable. Internal

filling solution will flow under

membrane and restore electrode

response


Low Slope or

standards contaminated

prepare fresh standards

No Slope or incorrectly made

buffer not used

use recommended buffer


standard used as buffer

use buffer


electrode exposed

hold electrode by outer body and

to air for extended period

pull on electrode cable. internal

filling solution will flow under

membrane

and

restore

electrode

response


membrane failure

replace membrane

(wet, perforation,

discoloration)


defective inner body

refer to

Checking the Electrode

Inner Body



"Incorrect Answer"

incorrect scaling

plot millivolts on the linear axis. On

(but calibration of semi-log paper

the log axis, be sure concentration

curve is good)

numbers within each decade are

increasing with increasing

concentration


incorrect sign

be sure to note sign of millivolt