SETTING UP YOUR THERMO DEVICES


For these devices:

91100-50 DuaLogR Thermocouple Thermometer
93210-50 ThermoLogR Thermister Thermometer
93410-50 ThermoLogR RTD Thermometer
37003-02 HumidityLogR Thermohygrometer


91100-50 DuaLogR Set-Up
The interval may be set to any time from once every 3 seconds to once
every hour. This example will set the thermometer to print live readings
once every 3 seconds.

1.

Press SETUP, then PRINT. The display indicates
“minutes:seconds” with the last printing interval setup shown.
The seconds will be flashing.

2.

Press MAX (up) or MIN (down), until the flashing digits indicate
“03” seconds.

3.

Press HOLD (enter). The display will show the minutes flashing.

4.

Press HOLD again. The large upper display will show “HP”, “300”
or “600”.

5.

Press MAX until “600” shows (use “300” for HYPERTERMINAL) in
the upper display.

6.

Press SETUP to exit.

Aim the top of the unit at the red window of the adapter at a distance of 12
inches or less.

93210-50 and 93410-50 ThermoLogR Set-Up
The interval may be set to any time from once every 3 seconds to once
every hour. This example will set the thermometer to print live readings
once every 3 seconds.

1.

Press SETUP, then PRINT. The display indicates
“minutes:seconds” with the

2.

last printing interval setup shown. The seconds will be flashing.

3.

Press MAX (up) or MIN (down), until the flashing digits indicate
“03” seconds.

4.

Press HOLD (enter). The display will show the minutes flashing.

5.

Press HOLD again. The large upper display will show “HP”, “300”,
“600”, “1200” or “2400”.

6.

Press MAX until “2400” shows in the upper display.

7.

Press SETUP to exit.

Aim the top of the unit at the red window of the adapter at a distance of 12
inches or less.

4