Wp phscan 3+ instructions, Ph testing, Reset to factory default – Thermo Fisher Scientific pHScan 3/3+ User Manual

Page 4: Temperature display, Before you begin, Calibration, Buffer set selection

background image

68X247747 11/00 Rev 0

4. Press CAL key to toggle between USA and NIST

buffer sets.

5. Press HOLD/CON key to confirm.

pH Testing

1. Remove cap from the electrode and press the

ON/OFF button to switch tester on.

2. Dip the electrode about 2 to 3 cm into the test

solution. Stir once and let the reading stabilize.

3. Note the pH or press HOLD/CON button to freeze

the reading. Press HOLD/CON again to release
the reading.

4. Press ON/OFF to turn off tester. If you do not

press a button for 8.5 minutes the tester will
automatically shut off to conserve batteries.

Reset To Factory Default

You can reset calibration data to factory default
settings. This is extremely useful if you want to take
a new set of pH measurements or using a new
replacement electrode.

1. With tester off, hold CAL key while pressing ON/

OFF key.

2. The display will show “rSt” flashing.

3. Press HOLD/CON key to confirm. The display

will flash “YES” for 2 seconds and the tester will
go back to measurement mode.

4. To exit without confirmation, press CAL key. The

display will flash “NO” for 2 seconds and the
tester will go back to measurement mode.

Temperature Display

The feature is extremely useful for you to note the
temperature of sample you are measuring.

1. With tester on, press HOLD/CON key. The display

will show “HO” momentarily.

2. Press CAL key.

3. Temperature value is displayed in °C for 5

seconds.

4. The tester will automatically go back to pH

measurement mode.

WP pHScan 3+ Instructions

Before You Begin

Remove electrode cap. To condition electrode,

immerse electrode in electrode solution, buffer

or tap water for at least 30 minutes. DO NOT use

deionized water

Calibration

Calibration should be done regularly, typically

everyday that the tester is used. You can calibrate

at up to three points (pH 4.01, 7.00 and 10.01

or 4.01, 6.86 and 9.18).

1. Press ON/OFF button to switch unit on.

2. Dip electrode about 2 to 3 cm into pH buffer

solution.

3. Press CAL button to enter Calibrate (CAL)

mode. ‘CAL’ flashes on the display. Then, a

pH value close to the pH buffer value will

flash repeatedly.

To abort calibration, press CAL again to

resume measuring mode.

4. After about 2 minutes when the

reading has stabilized, press the HOLD/

CON button to confirm calibration. The

display will show ‘CO’ and then switch

to the buffer value reading.

5. Repeat with other buffers if

necessary. Rinse electrode in tap

water before dipping into next buffer.

Insufficient sampling time. If
the tester does not have a long
enough exposure to the buffer, a
stable calibration point will not be
reached. Wait for about 2 minutes
when the reading has stabilized,

press the HOLD/CON.

Buffer Set Selection

To choose USA (4.01, 7.00, 10.01)

or NIST (4.01, 6.86, 9.18) buffers.

1. With tester off, hold HOLD/CON key

while pressing and releasing ON/OFF

key.

2. The display will show “bUF” flashing.

3. Press HOLD/CON key. The display

will show either “USA” or “NST”.