Wp phscan 3 instructions, Before you begin, Calibration – Thermo Fisher Scientific pHScan 3/3+ User Manual

Page 2: Buffer set selection, Ph testing, Reset to factory default

background image

Before You Begin

Remove electrode cap. To condition electrode,

immerse electrode in electrode solution, buffer

or tap water for at least 30 minutes. DO NOT use

deionized water

Calibration

Calibration should be done regularly, typically

everyday that the tester is used. You can calibrate

at up to three points (pH 4.01, 7.00 and 10.01

or 4.01, 6.86 and 9.18).

1. Press ON/OFF button to switch unit on.

2. Dip electrode about 2 to 3 cm into pH buffer

solution.

3. Press CAL button to enter Calibrate (CAL)

mode. ‘CAL’ flashes on the display. Then, a

pH value close to the pH buffer value will

flash repeatedly.

To abort calibration, press CAL again to

resume measuring mode.

4. After about 2 minutes when the

reading has stabilized, press the

HOLD/CON button to confirm

calibration. The display will show ‘CO’

and then switch to the buffer value

reading.

5. Repeat with other buffers if

necessary. Rinse electrode in tap
water before dipping into next buffer.

Insufficient sampling time. If
the tester does not have a long

enough exposure to the buffer, a stable

calibration point will not be reached.

Wait for about 2 minutes when the

reading has stabilized, press the

HOLD/CON.

Failure to re-hydrate the

electrode. A dry electrode will give

fluctuating readings while it

re-hydrates in a buffer, causing error.

68X247746 11/00 Rev 0

Buffer Set Selection

To choose USA (4.01, 7.00, 10.01) or NIST (4.01,
6.86, 9.18) buffers.

1. With tester off, hold HOLD/CON key while

pressing and releasing ON/OFF key.

2. The display will show “bUF” flashing.

3. Press HOLD/CON key. The display will show

either “USA” or “NST”.

4. Press CAL key to toggle between USA and NIST

buffer sets.

5. Press HOLD/CON key to confirm.

pH Testing

1. Remove cap from the electrode and press the

ON/OFF button to switch tester on.

2. Dip the electrode about 2 to 3 cm into the test

solution. Stir once and let the reading stabilize.

3. Note the pH or press HOLD/CON button to freeze

the reading. Press HOLD/CON again to release
the reading.

4. Press ON/OFF to turn off tester. If you do not

press a button for 8.5 minutes the tester will
automatically shut off to conserve batteries.

Reset To Factory Default

You can reset calibration data to factory default
settings. This is extremely useful when you want to
take a new set of pH measurements or when using
a new replacement electrode.

1. With tester off, hold CAL key while pressing ON/

OFF key.

2. The display will show “rSt” flashing.

3. Press HOLD/CON key to confirm. The display

will flash “YES” for 2 seconds and the tester will
go back to measurement mode.

4. To exit without confirmation, press CAL key. The

display will flash “NO” for 2 seconds and the
tester will go back to measurement mode.

WP pHScan 3 Instructions