RB-5A
Minimum temperature
-20
o
C
Cooling capacity at:
20
o
C
145 W
0
o
C
145 W
-10
o
C
110 W
Nominal dimensions and weights
Overall:
Length
430 mm
Width
250 mm
Height
566 mm
Bath opening:
Length
192 mm
Width
151 mm
Depth
200 mm
Maximum working capacity
7 litres
Net Weight
23 kg
RB-12A
-35
o
C
240 W
240 W
230 W
430 mm
370 mm
610 mm
208 mm
300 mm
150 mm
11.6 litres
44 kg
RB-22A
-30
o
C
240 W
240 W
230 W
430 mm
370 mm
690 mm
208 mm
300 mm
225 mm
22 litres
49 kg
* Including the Thermoregulator
Specification
The minimum temperatures quoted in the specification are
achieved in an ambient of 20
o
C with the appropriate liquid in the
integral bath. However many things will affect the minimum
obtainable temperature, such as: ambient temperature, insulation
and length of circulation system, condition of the working liquid,