DB200/2 AND DB200/3

Specification

DB200/2

DB200/2/115

DB200/3

DB200/3/115

Number of insert blocks

2

3

Temperature display

5 character LED

5 character LED

Working temperature range

Ambient +5°C to 200°C

Ambient +5°C to 200°C

Minimum set temperature

0°C

0°C

Set point resolution

0.1°C

0.1°C

Set point accuracy at 37°C

≤±1°C

≤±1°C

Set point accuracy at 100°C

≤±1°C

≤±1°C

Temperature stability at 37°C

≤±0.1°C

≤±0.1°C

Temperature stability at 100°C

±0.15°C

±0.15°C

Max. temperature variation within a block

<0.2°C at 37°C

<0.2°C at 100°C

<0.2°C at 37°C

<0.2°C at 100°C

Max. temperature variation between

similar blocks

<0.5°C at 37°C

<1.5°C at 100°C

<0.5°C at 37°C

<1.5°C at 100°C

Heat up time 25°C to 100°C

8 minutes

7 minutes

Heat up time 25°C to 200°C

23 minutes

19 minutes

Countdown Timer with audible beeps

Yes

Yes

Minimum / Maximum timer

1min / 99h59min

1min / 99h59min

Safety over-temp cut-out

Thermal fuse

Thermal fuse

Overall dimensions (W x D x H)

202 x 260 x 105mm

279 x 260 x 105mm

Electrical supply

230V or 115V, 50-60Hz

230V or 115V, 50-60Hz

Power

300W

450W

Shipping weight

5kg

6kg

Note: The control specifications are quoted for an ambient temperature of 20°C for units fitted with

2, 3 or 4 insert blocks type F3506. The specification will be closely held if the ambient temperature is

in the range 10°C to 30°C. Outside this range the quoted figures may deteriorate but the unit will still

work safely.

Note: Different block inserts may take different amounts of time to equilibrate to the same
temperature.