Ar ch ive d, Includes dates before 5/05) – Seametrics EX100-200 Series v.1 User Manual
Page 5

Page 3
INSTALLATION
Depth Setting. It is important for accuracy that the
sensor be inserted the correct depth into the pipe.
Follow these steps:
1. In Table 1, find the model of sensor which you have.
Take Dimension C from this table for the pipe size.
2. Subtract wall thickness of your pipe to find Dimension
D. Use Table 2 if you do not know the thickness.
3. Measuring from the outside of the pipe to the joint in
the housing, as shown in the diagram, adjust the sensor
to Dimension D and tighten enough to hold it there.
4. Line the conduit housing with the centerline of the
pipe, as shown.
5. Check Dimension D one more time.
6. Tighten the compression nut fully.
"D"
Caution!
Never attempt to remove a
flow sensor when there is pressure in
the pipe. Loosen the locking collar
slowly to release any trapped pressure. If fluid
sprays out when removing the sensor, stop turn-
ing and depressurize the pipe. Failure to do so
could result in the sensor being thrown from the
pipe, resulting in damage or serious injury.
PROPER DEPTH SETTING
AR
CH
IVE
D
(Includes Dates Before 5/05)