Martel Electronics LC100 User Manual
Page 14
![background image](/manuals/369799/14/background.png)
3) Press the
key once for the step
mode, twice for the slow ramp mode and
three times for the fast ramp mode. With
each press a small annunciator will appear
on the lower left side of the LCD indicating
which one of the step/ramp modes has been
selected.
4) Once enabled the step/ramp mode will run
continuously until another key is pressed.
B. Percent Error Mode
This unique feature will calculate error as a per-
centage of span (4 mA to 20 mA yields a 16 mA
span). The percent error feature allows the
user to “Zero Out” the %mA display so that the
deviation from the ideal value can be displayed
on the % error display. The procedure is as fol-
lows:
1) Connect the LC100 to the unit under test.
2) Place the LC100 in one of the mA output
modes (source or simulate).
3) Adjust the LC100 to the desired output.
4) If the unit under test does not read exactly as
it should, hold down the
key and
adjust the knob unit until the unit under test
reads correctly. Continue to hold down the
key.
5) The reading in the % display area will be the
error or deviation calculated as a percentage
of span.
%ERROR
%ERROR
AUTO