Advanced operating modes – Martel Electronics LC100 User Manual
Page 13
input impedance of the LC100
≥ 1 meg ohm
while in this mode.
Figure 6 - Measuring 0 to 28 VDC
5. Advanced Operating Modes
A. Auto Step/Ramp Mode
The LC100 has the ability to step the mA output
in 25% steps automatically in 5 second inter-
vals. It can also ramp the mA output from 4 mA
to 20 mA then back to 4 mA in a linear fashion
automatically.
To activate the Auto-Step/Ramp mode use the
following procedure:
1) The Auto Step/Ramp function can be used in
either the source mA or simulate mA mode.
Place the LC100 in one or the other mode.
2) Connect the LC100 to the device under test.
11
See also other documents in the category Martel Electronics Equipment:
- 3001 (94 pages)
- BetaGauge 301 (40 pages)
- BetaGauge 311A-Ex (32 pages)
- BetaGauge 311A (48 pages)
- BetaGauge 330 (48 pages)
- BetaGauge 3II (197 pages)
- BetaGauge PI (20 pages)
- BetaGauge PIR (24 pages)
- BetaLOG-PI (28 pages)
- BetaLOG-TI (16 pages)
- BetaProbe TI/TI+ (9 pages)
- DPC-30 (12 pages)
- DPC300A (48 pages)
- DMC1410 (108 pages)
- IVC-222HPII (12 pages)
- LC-110/H (50 pages)
- BETA LDT-500 (2 pages)
- M2000 (28 pages)
- MC1010 (46 pages)
- MC1210 (56 pages)
- TC100 (16 pages)
- T150 (8 pages)
- T140 Series (12 pages)
- PTC8010 (36 pages)
- PSC4010 (36 pages)
- MS-420 (1 page)
- MECP500 (2 pages)
- MECP2000 (2 pages)
- MECP10K (4 pages)
- MECP100 (2 pages)
- T130 (2 pages)
- T120 (2 pages)
- MC1200 (50 pages)
- MC1000 (28 pages)
- MasterCAL (140 pages)
- DMC1400 (106 pages)
- 320 (60 pages)
- 230 (106 pages)
- 130 (85 pages)
- QM100V (1 page)
- QM100V-40 (1 page)
- QM110T (2 pages)
- QM120ET (2 pages)
- QM130M (2 pages)