9. BetaGauge PIR Serial Interface

Instructions

Initiating Communication

The terminal communications can be setup using terminal
communication software on a PC. The terminal settings need to
be set as follows:

• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1
• Flow control: None
• Local echo on

List of Commands

Command

Description

CAL_START

Puts the calibrator in calibration mode

*CLS

Clears the error queue.

FAULT?

Returns an error code from the error queue

*IDN?

Identification query. Returns the manufacturer, model
number, and firmware revision level of the Calibrator.

TARE

Tares the offset pressure of the reading on the calibra-
tor

TARE?

Returns the current tare value

PRES_UNIT?

Returns the pressure unit for the upper display.

PRES_UNIT

Sets the pressure unit for the display

ZERO_MEAS

Zeros pressure of the calibrator

ZERO_MEAS?

Returns the current zero offset value

MINMAX_RST

Resets the minimum and maximum recorded values.

MIN?

Returns the minimum recorded value

MAX?

Returns the maximum recorded value

TEMP?

Returns temperature in the chosen units

CAL_STORE

Stores calibration data.

CUST_MULT?

Sets the multiplier for the custom unit type

STREAM_OFF

Turns streaming data off

STREAM_ON

Turns streaming data on

VAL?

Returns the measured pressure value in selected units

TEMP_UNIT

Used to set temperature unit

TEMP_UNIT?

Returns temperature unit

15

This manual is related to the following products: